Key Features:
Dual Cell High and Low Temperature Consistometer – Equipped with two independent testing cells, this instrument allows simultaneous measurements at different temperatures, improving efficiency and comparative analysis.
Double Cylinders Large Temperature Difference Range Consistometer – The dual-cylinder design ensures stable performance across a wide temperature range, enabling testing under extreme conditions (e.g., from sub-zero to high-temperature environments).
Precision Control – Advanced temperature regulation and real-time data acquisition ensure accurate and repeatable results.
User-Friendly Interface – An intuitive software system provides easy operation, data logging, and analysis.
Maximum Temperature | 250 ℃ / 482 ℉ |
Maximum Pressure | 200 MPa / 29,000 psi |
Minimum Temperature | Lower than 0℃ / Lower than 32℉ |
Max. Pressure under Min. Temp | 150 MPa / 21,760 psi |
Maximum Heating Rate | 3℃/min / 37℉/min |
Average Chilling Rate | 2.5℃/min / 36.5℉/min |
Temperature Precision | ±1℃ / ±1.8℉ |
Pressure Precision | ±0.7 MPa / ±102 psi |
Speed Range | 0 to 600 rpm |
Power Source | 380 volts 50Hz three-phrase |
Control Unit Power | 4,200 watts (×2) |
Chiller Power | 4,000 watts (×2) |
Speed Control Power | 1,000 watts |
Compressed Air | 0.4 to 0.7 MPa / 58 to 102 psi |
Cooling Water | 0.2 to 0.6 MPa / 29 to 87 psi |
Dimensions | 1,300×740×1,75 0mm3 / 50×30×70 in3 |
Weight | 1,000kg / 2,205 lb |
Evaluation of drilling fluid rheology under simulated wellbore conditions
Testing cement slurry consistency in oilfield applications
Quality control in chemical and material manufacturing